| Emerson's OPM
3000 Opacity Dust Density Monitor ratio measurement technique provides continuous automatic compensation for variations inlight
source intensity to ensure prolonged instrument accuracy and stability. Since the OPM
3000 is insensitive to the absolute intensity of the light source, it is not
affected by light source aging. |
Emerson's
Rosemount Analytical OPM 3000 stack
mounted transmissometer sensor system consists
of an optical transceiver mounted on one side of the
stack and a retro reflector mounted on the other.
The main light source is electronically modulated
and projects a collimated beam of light, which is
split into a reference beam and a measurement
beam. The measurement beam is projected across
the stack to a retro reflector, which reflects it back
across the stack to the measurement detector. The
output of the measurement and reference detectors
are transmitted to the control room mounted monitoring
unit where all output signals are analyzed.
|
 |
|